Veterinary Sterilization Laboratory Autoclave With Visually And Audibly Alarm
Brief
This Autoclave Steam Sterilizer is designed to sterilize veterinary instruments from veterinary
laboratories, veterinary quarantine station, epidemic prevention station , livestock farm and animal
hospitals etc.
Designed and manufactured according to ISO 9001, ISO 13485 system and latest moist steam
sterilizer related standards, which has the obvious advantage of low failure rate and easily
operation to ensure your satisfied using smoothly.
Safety
Door Safety
• Chamber door cannot be opened when chamber is pressurized
• Steam cannot enter the chamber when the door is open
• A cycle cannot start if the door is open or not properly locked
• Sliding door movement is stopped if an object is detected in front of the door
Safety Valves
• Both chamber and jacket are equipped with pressure relief valves to prevent excess pressure.
• Saturated steam can be supplied to the autoclave from building steam equipped safety valve
• A stand-alone generator or a built-in steam generator with safety valve is available for your option
Emergency Shut-Off
• Easily accessible emergency switches for immediate cycle
shut-off.
• Double Independent Monitoring
• Digital and mechanical monitoring systems provide accurate pressure results which are displayed
on the analog and digital displays.
Sterilization Programs
1, B-D Test
3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min.
2, Vacuum Leak Test
Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect.
3, Instruments/Tools
3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min.
4, Clothing
3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min.
5, Liquid
15 min exchange, expose temperature 121℃, expose time 15 min, cooling time 25 min.
6, User Programmable
All preset parameters can be resettable; total 15 kinds programs can be customized
Loading/Unloading System
Standards equipped: 1 piece loading cart of SUS316L stainless steel, 1 piece trolley of SUS304
stainless steel for single door sterilizer and 2 pieces trolley for double door.
Loading plate, racks or shelves are available for option.
All loading/unloading system features of,
• High capacity for maximum throughput
• Eliminates operator handling of hot loads
• Swivel casters for easy maneuverability
• Low profie docking mechanism for easy alignment
• Racks and shelves for selected chamber sizes
Main Parameters
Model | MD-0.6V |
Door type | Vertical sliding single/double door |
Max working pressure | 0.23Mpa |
Working vacuum degree | <-0.086Mpa |
Max working temperature | 134℃ |
Chamber size | 1670*600*600 mm(L*W*H) |
External dimension | 2010*1350*1800mm(L*W*H) |
Net weight | 1200Kg |
Utilities Supply | |
Electric power | AC380V, 50Hz, 3Ph/N/E,3KVA |
Water | 0.15---0.3MPa, Hardness <50ppm, pH Neutral, 55Kg/cycle |
Steam | 0.5~0.7Mpa, 30Kg/cycle |
Compressed air | 0.5~0.7Mpa, 10 liters/minute |
Medpharm laboratory autoclave
Autoclave alarm
ZHANGJIAGANG MEDPHARM MACHINERY LTD.
MEDPHARM Company team is committed to providing quality sterilization products, service, parts and
maintenance training at significant savings. From customized parts packages to turnkey equipment
acquisition and installation, our innovative sterilization solutions are designed to meet our customers' needs.
We understand that untimely breakdowns are costly and can be detrimental to your everyday operations.
Our service team is compromised of expert technicians that respond quickly to all of your service needs.
We offer on-site service support including inservices, installation and deinstallation of your equipment. From
breakdown repairs to comprehensive maintenance programs, we have an innovative service solution for you.
We will service on-Site Repairs.Our staff is equipped with the resources required to provide you with
on-site, on-time repairs in compliance with manufacturer specifications. We also provide installation,
deinstallation, and inservices on most makes and models of sterilization equipment. Near or far we can
provide you with service support that you need.
Our reconditioning background has taught us of the importance of having quality parts on hand at all times.
We have a large inventory of individual repair parts, as well as preventive maintenance kits, and consumables
in stock and ready to ship today.
With our expert technical support staff, it is not always necessary to have part numbers to place an order.
We will help you identify the quality parts that you need.