Strip Chart Recorder Autoclave Sterilizer Machine With Fault Identifications Incorportated
Specifications
1. Sanitary stainless steel piping
2. Cross-contamination or gas-tight barrier (Bioseal)
3. Own control system, specially developed for sterilization applications (choice of world-leading PLC brands)
4. Manufactured according to intended application guidelines/standards and country of installation
5. Project handling and products in line with current regulations and guidelines, e.g. cGMP, etc.
Program Parameter
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: ●Packaged instruments ●Empty glass containers ●Empty metal containers Porous materials to be sterilized: ●Rubber articles |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced ●Textile |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test programs | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leak Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |
Description
Autoclave sterilizer machine sterilization relies on three (3) parameters to eliminate microbes
and organisms: Time, temperature, and pressure. And these parameters can be manipulated
into different cycle recipes to sterilize various types of loads. However, the cycle requirement for
every load type—such as red bags, media, glassware, cages, animal bedding, and pipette
tips—can vary significantly.
While the vast majority of common pharmacy and laboratory loads can be sterilized with one of
the following three (3) basic cycles: Gravity, Vacuum (or Prevac), and Liquids; it is important to
ensure that your sterilizer offers the cycles appropriate for all of your load requirements.
Clients will be of particular interest to labs that are expanding their research capabilities to mice
and rats, and need to upgrade their current sterilizer to one that can easily accommodate
cages and waste.
Some of the key available options include:
Pass-thru configuration: Ideal for BSL-3 labs, the pass-thru option allows for movement of
materials between the contained decontamination area and the clean processing area.
Loading cart and transfer carriage: The loading cart is mounted on rollers, with two stainless steel
shelves and a wheeled transfer carriage. The transfer carriage mates to interior guide rails to
ease loading and unloading of sterilization loads up to 500 lbs. evenly distributed.
Bioseal: For Biological Safety Level 3 installations a Bioseal can be installed to maintain the
contamination integrity of the clean side. Utilities can be mounted at the non-contained side
of the installation and in the front, rear or designated side of the cabinet.
Dual Controls: For pass-thru models equipped with the controller. A 6” color touchscreen with
full operator and service functionality is provided at each end of the sterilizer to allow operation
from both sides. Printer is located outside the containment area unless otherwise specified.
Applications
BioPharma Series Sterilizers provide steam sterilization for components and equipment as well
as for container closures such as rubber stoppers and plungers which are vital to parenteral packaging.
Programming for many applications including stopper processing, which generally clean, sterility and drying of these products that are a functional part of the drug administration system.
Advantage
ZHANGJIAGANG MEDPHARM MACHINERY LTD.
MEDPHARM Company team is committed to providing quality sterilization products, service, parts and
maintenance training at significant savings. From customized parts packages to turnkey equipment
acquisition and installation, our innovative sterilization solutions are designed to meet our customers' needs.
We understand that untimely breakdowns are costly and can be detrimental to your everyday operations.
Our service team is compromised of expert technicians that respond quickly to all of your service needs.
We offer on-site service support including inservices, installation and deinstallation of your equipment. From
breakdown repairs to comprehensive maintenance programs, we have an innovative service solution for you.
We will service on-Site Repairs.Our staff is equipped with the resources required to provide you with
on-site, on-time repairs in compliance with manufacturer specifications. We also provide installation,
deinstallation, and inservices on most makes and models of sterilization equipment. Near or far we can
provide you with service support that you need.
Our reconditioning background has taught us of the importance of having quality parts on hand at all times.
We have a large inventory of individual repair parts, as well as preventive maintenance kits, and consumables
in stock and ready to ship today.
With our expert technical support staff, it is not always necessary to have part numbers to place an order.
We will help you identify the quality parts that you need.